Temperature Control System for Cell Continuous Observation under the Microscope |
LI Yuan, PENG Ming-qing, GONG Fang, DIAO Qi-zhi, LIU Bei-zhong |
Central Laboratory of Yongchuan Hospital, Chongqing Medical University, Chongqing 400016, China |
|
|
Abstract This paper developes a temperature control system in order to carry out a real time and continuous observation on living cells under the microscope. In this system, transparent indium tin oxid(ITO) conductive film was used as a heater, temperature was measured by the temperature dependence of resistance behavior of the ITO film, C8051F340 single-clup was used as the control hardware core, and the modified PID algorithm and pulse width modulation (PWM) were adopted as the key control software programs. The temperature system had a simple structure without added temperature sensor, an uniform temperature distribution within ±1℃ on the ITO film surface and temperature control precision of ±0.2℃ could be obtained. The system could maintain a good growth state for BHK-21 living cell under the microscope for 48 h.
|
Received: 10 January 2016
|
|
Corresponding Authors:
PENG Ming-qing. E-mail: mingqingqing@yahoo.com.cn
|
|
|
|
[1] Yang Caibiao, Chen Qiang, Shao Jianbo, et al. Study on glass cell culture chips[J]. Letters in Biotechnology, 2008, 19(1): 76-79. [2] Yang Caibiao, Zhao Jianlong. Chip temperature control system based on 812[J]. Microcomputer Information, 2007, 23(25):43-44. [3] Hung PJ, Lee PJ, Sabounchi P, et al. A novel high aspect ratio microfluidic design to provide a stable and uniform microenvironment for cell growth in a high throughput mammaliancell culture array[J]. Lab on a Chip, 2005, 5(1): 44-48. [4] Dai Jing, Fan Xiaofeng, Fang Jin, et al. Study on temperature measurement and control for microfluidic systems[J]. Spectroscopy and Spectral Analysis, 2008, 28(1):145-152. [5] Wu Zhiyong, Tian Xiaoxi, Qu Baiyan, et al. Self-heating and sensing static chip polymerase chain reaction(PCR) of transparent electroconductive glass substrate(ITO)[J]. Chemical Journal of Chinese University, 2007, 28(12): 2259-2263. [6] Zhu Lei. Design of the low cost data acquisition device based on C8051F340[J]. Internatioanl Electronic Elements, 2008, 4:6-8. [7] Zhao Xiaoqing, He Guangping, Yuan Junjie, et al. A design of temperature controller for rapid thermal processing[J]. Application of Electronic Technique, 2010, 36(2):133-136. [8] Sun K, Yamaguchi A, Ishida U, et al. A heater-integrated transparent microchannel chip for continuous-flow PCR[J]. Sensors and Actuators B: Chemica, 2002, 84(2-3): 283-289. |
|
|
|